Add products by adding codes
- Original MELAG ME34150 chamber door gasket for autoclaves MELAG Typ 23/24/28/29/30 and MELAtronic 23N/23N+. Ensures tightness and safe sterilization. Ideal for medical, dental and beauty practices.
CHAMBER DOOR GASKET MELAG TYPE 23 / MELATRONIC 23 / 23EN (ME34150)
Original SILICONE door gasket for MELAG TYPE 23 / MELATRONIC 23 / 23EN autoclaves. Ensures reliable chamber sealing, stable pressure and temperature parameters, and safe sterilization cycles. Recommended as a replacement element in case of wear, deformation or steam leaks around the door.
The gasket is manufactured to MELAG factory specifications, providing precise fit to the chamber door and long service life under repeated thermal loads. Replacement can typically be performed during routine maintenance without additional modifications to the unit.
Product information
- Manufacturer: MELAG
- Product type: chamber / door gasket
- Application: sealing of chamber door in MELAG TYPE 23 and MELATRONIC 23 / 23EN autoclaves
- Material: high‑temperature SILICONE
- Colour: white
- Original manufacturer code: ME34150
- Article type: spare part / consumable
- Installation: manual replacement during service or maintenance
Compatibility
In most spare parts and consumables, compatibility with other devices and technical parameters are listed below the product description.
This gasket is compatible with the following MELAG autoclaves:
- MELAG TYP 23
- MELATRONIC 23
- MELATRONIC 23EN
- MELATRONIC 24
- MELATRONIC 28
- MELATRONIC 29
- MELATRONIC 30
- MT23
Not suitable for MELAG PREMIUM, PROFI or EUROKLAV (S‑CLASS) series autoclaves.
Technical and service notes
- Check the door sealing surface for contamination or mechanical damage before installation.
- It is recommended to replace the gasket when visible cracks, hardening or permanent deformation are observed, or when steam leaks are detected during the sterilization cycle.
- After replacement, perform a vacuum and leak test (if available in the device) before returning the autoclave to regular operation.
- Use only original MELAG gaskets (code ME34150) to maintain device safety and conformity with manufacturer’s specifications.